Brookeld Engineering Labs., Inc. Page 16 Manual No. M08-372-C0113
III.6 Operating Menu
The CT3 Texture Analyzer offers seven test modes, one calibration mode and one parameter selec-
tion table.
Normal – single compression cycle.
Hold Time – compress and hold.
Cycle Count – compress sample several times.
Bloom – compression test measures gelatin bloom strength.
TPA–twocycleTextureProleAnalysiscompressiontest.
Tension – pulls apart a sample using tensile force.
Surimi – single compression cycle, stops automatically after gel breaks.
Static Load – monitors load while hanging weights for calibration.
Global Parameter – allows setting pretest speed, post test speed and temperature units.
Each test mode requires parameters to be set. All test mode selection and parameter setting is done
usingtheSelect/Scrollknob(seegure1.2).Thisknobisbothrotatedanddepressedduringopera-
tion.
NOTE:Whensettingthetestspeed,pressingtheRESETbuttonwilltoggletheincrementsfrom
0.1mm/s to 1mm/s.
Thespecicparametersrequiredwilldependuponthetypeoftestchosen:
NormalTestRequires: trigger,deformation,speed
HoldTimeTestRequires: holdtime,trigger,deformation,speed
CycleCountTestRequires: cyclecount,trigger,deformation,speed
BloomTestRequires: Allparametersarexedaccordingtoindustrystandard
TPATestRequires: trigger,deformation,speed
TensionTestRequires: trigger,deformation,speed
SurimiTestRequires: trigger,correction,speed
StaticLoadTestRequires: none
GlobalParamters: pretestspeed,posttestspeed,temperatureunits
These parameters, once set, will be maintained by the CT3 even after power down to facilitate re-
petitive testing. The test mode used prior to powering down will be presented on the screen at the
next power up. A test is started by depressing the start button.
Thetestparametersaredenedasfollows:
Trigger: The load, in grams, measured by the CT3 to indicate that the probe is in contact
withthesample.Oncethetriggervalueisreached,thetestwillbeginatthedened
speed.BrookeldrecommendstriggervaluesasspeciedinTableII.1Setting
trigger value to zero disables trigger function so test will immediately start
from current position.
Deformation: The total downward distance the probe will travel once the trigger value is
reached.
Speed: Thespeedatwhichtheprobewilltravelthespecieddistance.
Time: Thenumberofsecondsthattheprobewillbeheldatthedeneddistanceduring
a Hold Time test.
Count: The number of cycles (Speed and Distance) that will be applied to the sample
during a Cycle Count test.
Correction: Speciesthesizeoftheloaddrop,ingrams,thatisrequiredtorecognizeagel
rupture.