Brookeld Engineering Labs., Inc. Page 1 Manual No. M02-312A0607BROOKFIELD CAP 1000+ VISCOMETERMODEL CAP 1000+Operating InstructionsManual No. M02-
Brookeld Engineering Labs., Inc. Page 8 Manual No. M02-312A0607 I.8 CleaningInstrument, Keypad & Painted Surfaces: Clean with dry, non-abras
Brookeld Engineering Labs., Inc. Page 9 Manual No. M02-312A0607 II. GETTING STARTED II.1 Power ONTurn the power ON using the switch located on th
Brookeld Engineering Labs., Inc. Page 10 Manual No. M02-312A0607When using the solvent trap (Figure II-2), connect it to the cone adapter by slidin
Brookeld Engineering Labs., Inc. Page 11 Manual No. M02-312A0607Note: 1. CAP Viscometers allow for only one cone at a time of the same cone num
Brookeld Engineering Labs., Inc. Page 12 Manual No. M02-312A0607It is recommended to have the spindle in contact with the plate prior to introducin
Brookeld Engineering Labs., Inc. Page 13 Manual No. M02-312A0607A maximum of 999 seconds can be printed when running in manual TIMER mode (00). Ov
Brookeld Engineering Labs., Inc. Page 14 Manual No. M02-312A0607 III. OPERATIONThe CAP 1000+ Viscometer is delivered from the factory with two av
Brookeld Engineering Labs., Inc. Page 15 Manual No. M02-312A0607Table III-2: Full Scale Range Viscosity for High Torque CAP 1000+Cone NumberCone R
Brookeld Engineering Labs., Inc. Page 16 Manual No. M02-312A0607Table III-3: Accuracy for Low Torque CAP 1000+Cone ≤400 RPM 750 RPM 900 RPM 900 RP
Brookeld Engineering Labs., Inc. Page 17 Manual No. M02-312A0607• In ambient conditions outside of 15°C to 30°C, accuracy is +/- 2.0°C when the te
ADDENDUMBrookeld CAP Viscometer Operating InstructionsCAP1000+ Manual No. M02-312A0607CAP2000+ Manual No. M02-313-F0211 Addendum No. D12-3312The ta
Brookeld Engineering Labs., Inc. Page 18 Manual No. M02-312A0607 Figure III-1a Figure III-1b Figure III-1c Figure III-1d 4. Run a vi
Brookeld Engineering Labs., Inc. Page 19 Manual No. M02-312A0607Table III-5: Viscosity Standard Fluids for Calibration of CAP Spindles on Low Torq
Brookeld Engineering Labs., Inc. Page 20 Manual No. M02-312A0607 III.4 Cone CalibrationA special feature of the CAP Series Viscometers allows the
Brookeld Engineering Labs., Inc. Page 21 Manual No. M02-312A0607 Note: The calibration process may be cancelled at any time prior to pressing RUN
Brookeld Engineering Labs., Inc. Page 22 Manual No. M02-312A0607 III.6 Making Viscosity MeasurementsThe following procedure is recommended for mak
Brookeld Engineering Labs., Inc. Page 23 Manual No. M02-312A0607 Notes: a. Lower the handle gently. DO NOT FORCE THE CONE ONTO THE PLATE.
Brookeld Engineering Labs., Inc. Page 24 Manual No. M02-312A0607 APPENDIX A - Variables in Viscosity MeasurementsAs with any instrument measurement
Brookeld Engineering Labs., Inc. Page 25 Manual No. M02-312A0607 Viscometer Related Variables• Most uid viscosities are found to be non-Newtonian
Brookeld Engineering Labs., Inc. Page 26 Manual No. M02-312A0607 APPENDIX B - Warranty Repair and ServiceWarrantyBrookeld Viscometers are guaran
Brookeld Engineering Labs., Inc. Page 27 Manual No. M02-312A0607 PACKAGING INSTRUCTIONS TO RETURN A BROOKFIELD CAP VISCOMETER FOR REPAIR OR CALIBRA
Table III-3 Accuracy for LOW TORQUE CAP ViscometerCone Number ≤ 500 RPM10-100% FSR01 ± 2%02 ± 2%03 ± 2%04 ± 2%05 ± 2%06 ± 2%07 ± 2%08 ± 2%09 ± 2%1
Brookeld Engineering Labs., Inc. Page 28 Manual No. M02-312A0607IMPORTANT: 1. Brookfi eld recommends that all viscometers be returned for annual ca
Brookeld Engineering Labs., Inc. Page 2 Manual No. M02-312A0607TABLE OF CONTENTS I. INTRODUCTION ...
Brookeld Engineering Labs., Inc. Page 3 Manual No. M02-312A0607 I. INTRODUCTIONThe CAP 1000+ Series Viscometers are high shear rate instruments w
Brookeld Engineering Labs., Inc. Page 4 Manual No. M02-312A0607 The following optional items may have been included: Part No. 7. Viscosit
Brookeld Engineering Labs., Inc. Page 5 Manual No. M02-312A0607 I.3 Specications Torque Range: Low 797-7,970 dyne•cm (designated on serial ta
Brookeld Engineering Labs., Inc. Page 6 Manual No. M02-312A06073) Verify that the viscometer’s power requirements match your power source BEFORE c
Brookeld Engineering Labs., Inc. Page 7 Manual No. M02-312A0607Figure I-3 I.6 Key Functions Figure I-3 shows the control keys on the face of the