Brookfield CAP1000 Viscometer Manual do Utilizador Página 18

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 30
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 17
Brookeld Engineering Labs., Inc. Page 16 Manual No. M02-312A0607
Table III-3: Accuracy for Low Torque CAP 1000+
Cone ≤400 RPM 750 RPM 900 RPM 900 RPM
10-100% FSR 10-100% FSR 50% FSR >50% FSR
01 ±2.0% ±2.0% ±2.0% ±2.0%
02 ±2.0% ±2.0% ±2.0% ±2.0%
03
±2.0% ±2.0%
±2.0%
±2.0%
04 ±2.0% ±2.0% ±2.0% ±2.0%
05 ±2.0% ±2.0% ±2.0% ±2.0%
06 ±2.0% ±2.0% ±2.0% ±2.0%
07 ±2.0% * * *
08 ±2.0% ±2.0% ±2.0% ±2.0%
09 ±2.0% ±2.0% ±2.0% ±2.0%
10 ±2.0% ±2.0% ±2.0% ±2.0%
* Not recommended for use at these RPM
Table III-4: Accuracy for High Torque CAP 1000+
Cone ≤400 RPM 750 RPM 900 RPM 900 RPM
10-100% FSR 10-100% FSR 50% FSR >50% FSR
01 ±2.0% ±2.0% ±2.0% ±4.0%
02 ±2.0% ±2.0% ±2.0% ±4.0%
03
±2.0% ±2.0%
±2.0%
±4.0%
04 ±2.0% ±3.0% ±3.0% ±6.0%
05 ±2.0% ±4.0% ±4.0% ±8.0%
06 ±2.0% ±5.0% ±5.0% ±10.0%
07 ±2.0% * * *
08 ±2.0% * * *
09 ±2.0% * * *
10 ±2.0% ±2.0% ±2.0% ±2.0%
* Not recommended for use at these RPM
The specication of temperature accuracy on CAP Viscometers is stated below:
LOW TEMP UNITS:
In ambient conditions of 15°C to 30°C, accuracy is +/- 0.5°C when the temperature set point
on the viscometer is 5°C to 75°C.
In ambient conditions outside of 15°C to 30°C, accuracy is +/- 1.0°C when the temperature
set point on the viscometer is 5°C to 75°C.
HIGH TEMP UNITS:
In ambient conditions of 15°C to 30°C, accuracy is +/- 0.5°C when the temperature set point
on the viscometer is 50°C to ≤100°C.
In ambient conditions of 15°C to 30°C, accuracy is +/- 1.0°C when the temperature set point
on the viscometer is 101°C to 235°C.
Vista de página 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 29 30

Comentários a estes Manuais

Sem comentários