Brookeld Engineering Labs., Inc. Page 6 Manual No. M11-353-A0512
Wire Cleaning Brush for Sample Tube FB53 1
Brush to clean loose debris from Falling Balls FB52 1
Leather cloth for polishing Falling Balls FB70 1
Sealant ring (perbunane) A 16x20 FB31 4
Sealant ring (silicone) 10x14x2 FB32 1
Thermometer, 0°C to +100°C* TM1 1
Operator Manual M11-353 1
Instrument Case FB200 1
Latex rubber tubing, 5/16 I.D. x 1/16 wall FB69 1
Certicate of Calibration ____ 1
*Other temperature measurement options are available. Contact your Brookeld dealer for
information.
I.2 Specifications
Viscosity Range: 0.5 - 7x10
4
mPa•s (cP)
Falling Time-Lower Limit: 60 s for Ball No.1
30 s for Ball Nos. 2,3,4,5 and 6
Falling Time-Upper limit: 300 s
Materials with viscosity >7x10
4
mPa•s require running
times of over 300 s.
Measuring Distance: 100 mm (50 mm between adjacent ring marks) in both
directions
Fall Tube Inner Diameter: 15.94mm +/- 0.01mm
Set of Balls: 6 balls
Working Angle: KF30: 80° (DIN position)
KF40: 80°, 70°, 60°, 50°
Temperature Range: -60°C - +150°C
Sample Volume: 40 mL
Dimensions: 180mm x 220mm x 330mm
Weight: 6.4 lbs, 2.9 kg
(empty glass tube and empty waterbath jacket)
Comentários a estes Manuais