- 41 -
Appendix G - Communications
When using the Brookfield Computer Cable (Brookfield part # DVP-80), the DV-II+ will output a
data string at a rate of approximately 2.5 times per second. When using the Brookfield Printer Cable
(Brookfield Part No. DVP-81), the output rate is 1.0 times per second. The DV-II+ uses the follow-
ing RS-232 parameters to output these strings:
Baud Rate 9600
Data Bits 8
Stop Bits 1
Parity None
Handshake None
The following formulas should be used to calculate and display the Viscometer data after each
packet of data is obtained from the DV-II+.
Viscosity (cP) = 100
*
TK
*
SMC
*
Torque
RPM
Shear Rate (
1
/
Sec
)=RPM
*
SRC
Shear Stress (
Dynes
/
Cm
2
)=TK
*
SMC
*
SRC
*
Torque
Where:
RPM = Current Viscometer spindle speed in RPM
TK = Viscometer torque constant from Appendix D, Table D2.
SMC = Current spindle multiplier constant from Appendix D, Table D1.
SRC = Current spindle shear rate constant from Appendix D, Table D1.
Torque = Current Viscometer torque (%) expressed as a number between 0 and 100.
As an example, consider an LV Viscometer using an SC4-31 spindle, running at 30 RPM and
currently returning a Torque of 62.3 (%). First we list all of the given data and include model and
spindle constants from Appendix D, Tables D1 and D2:
RPM = 30 from the example statement.
TK = 0.09373 from Appendix D for an LV Viscometer.
SMC = 32.0 from Appendix D for a type SC4-31 spindle.
SRC = 0.34 from Appendix D for a type SC4-31 spindle.
Torque = 62.3 from the example statement.
Applying this data to the above equations yields:
Viscosity (cP) =
100
*
TK
*
SMC
*
Torque
RPM
= 100 * 0.09373 * 32.0 * 62.3
30
= 622.9 cP
Shear Rate (
1
/
Sec
)=RPM
*
SRC
= 30 * 0.34
= 10.2
1
/
Sec
Comentários a estes Manuais