BrookeldEngineeringLabs.,Inc. Page58 ManualNo.M08-218-B0212
StepNr1
Start
[‰] :
250
End
[‰] :
250
Nr. Of Mp: 40
time [s] : 60
StepNr2
Start
[‰] :
500
End
[‰] :
500
Nr. Of Mp: 40
time [s] : 60
StepNr3
Start
[‰] :
750
End
[‰] :
750
Nr. Of Mp: 40
time [s] : 60
6. Enter the following information in the general block data:
a. Viscosity Standard Name (e.g. B11000)
b. Lot Number
c. Expiration Date
e. Viscosity Value
f. Spindle Serial Number
7. Upon test completion, click the Analysis button. The average viscosity of each step will ap-
pear in the bottom window.
8. An average viscosity reading should be within ±3% of the standard’s stated viscosity value at
the appropriate temperature.
B.5.3Troubleshooting
Check each of the items below if your measured values fall outside the range of uncertainty:
a. Assure that the gap is set properly.
b. Assure that the temperature is set properly.
c. Assure that correct amount of calibration uid is present and the gap is lled properly.
d. Assure that correct spindle constants are being used.
e. Assure that the rheometer has been zero calibrated.
Error Messages
Message Cause Solution
Range Error User input a value less than
the allowed minimum value
or greater than the allowed
maximum value (Speed or
Torque)
Check values. The R/S will advise
the maximum and minimum values
when this error is displayed.
Zero Cal. Error! An unacceptable value was
measured during the zero
point calibration.
Ensure the spindle is not mounted to
the R/S. Re-try zero calibration. If
this fails, call for service.
ABORT: Speed Max! Fluid is too thin to be run
at requested Shear Stress/
Torque level. Alternatively, no
uid is present.
This should not happen following the
proceding procedure.