Brookeld Engineering Laboratories, Inc. Page 60 Manual No. M/07-022-D0613
Calibration Check Procedure for Cone/Plate Viscometers
1) Follow the procedures outlined in Appendix A for mechanically adjusting the setting of
the cone to the plate.
2) Refer to Appendix A, Table A-1, and determine the correct sample volume required for
the spindle to be utilized.
3) Select a viscosity standard uid that will give viscosity readings between 10% and 100%
of Full Scale Range. Refer to Appendix B for viscosity ranges of cone spindles. Do not
use a silicone viscosity standard uid with a viscosity value greater than 5000 cP with
a Cone/Plate Viscometer. Brookeld offers a complete range of mineral oil viscosity
standards suitable for use with Cone/Plate Viscometers. See Table E-2 in Appendix E.
It is best to use a viscosity standard uid that will be close to the maximum viscosity for
a given cone spindle/speed combination.
LVDV-I PRIME Viscometer, Cone CP-42, Fluid 10
Having a viscosity of 9.7 cP at 25°C
At 60 RPM, the Full Scale Viscosity Range is 10.0 cP. Thus, the Viscometer reading
should be 97% torque and 9.7 cP viscosity ± 0.197 (closer to ±0.2) cP. The accuracy is
a combination of Viscometer and uid tolerance (refer to Interpretation of Calibration
Test Results at the end of this section).
4) With the viscometer stopped, remove the sample cup and place the viscosity standard
uid into the cup.
5) Connect the sample cup to the Viscometer. Allow sufcient time for temperature to reach
equilibrium.
6) Measure the viscosity and record the Viscometer reading in both % torque and centipoise.
NOTE: The spindle must rotate at least ve (5) times before readings are taken.