Brookfield DV3T Rheometer Manual do Utilizador Página 119

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 125
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 118
Brookeld Engineering Labs., Inc. Page 119 Manual No. M13-2100
Appendix I - DVE-50A Probe Clip
ProbeClipDVE-50Ais suppliedwithallmodelDV3TRheometers,DV-IIIRheometers,and
DigitalTemperatureIndicators.ItisusedtoattachtheRTDtemperatureprobetotheLVGuard
Leg(PartNo.B-20Y)or600mLlowformGrifnbeaker.FigureI-1isaviewoftheProbeClip,
showingtheholeintowhichtheRTDprobeisinserted,andtheslotwhichtsontotheLVguard
leg.WheninsertingtheRTDprobeintotheProbeClip,theupperpartoftheClipiscompressed
bysqueezingthepointsshowninFigureI-1.
Note: AllViscometer/Rheometermodels—exceptLV—usetheProbe
ClipasshowninFigureI-3.
RTD Probe
Hole
Squeeze Here when
installing RTD
Temperature probe
Figure I-1
FigureI-2shows the ProbeClip(withRTD temperature probe installed)mountedonthe LV
guard leg.
FigureI-3showstheProbeClipmountedina600mLlowformGrifnbeaker.Thismounting
maybeusedwithLV,RV,HAandHBseriesinstruments.
Caution: Temperature probe must not contact the spindle during
measurement.
Note: TheRTDprobemustbeparalleltothebeakerwallsoasnotto
interferewiththeviscositymeasurement.
Figure I-2
Figure I-3
Brookfield 600 mL Low Form
Griffin Beaker
DVE–50
Probe Clip
RTD Temperature Probe
LV Guard Leg
Assembly B-20KY
Probe Clip
DVE-50
RTD Temperatur e
Probe
Vista de página 118

Comentários a estes Manuais

Sem comentários