Brookeld Engineering Labs., Inc. Page 121 Manual No. M13-2100
❏ Rheometer Will Not Return to Zero
✓ Rheometerisnotlevel
• Checkwithspindleoutofthesample.
• Adjustthelaboratorystand.
✓ Pivotpointorjewelbearingfaulty
• PerformanOscillationCheck*
✓ RemovethespindleandturnthemotorOFF;selectdisplayto%torquemode.
✓ Gentlypushupontherheometercoupling.
✓ Turnthecouplinguntilthedigitaldisplayreads10-15onthe%display.
✓ Gentlyletgoofthecoupling.
✓ Watchthedigitaldisplay;youshouldseea“run”ofnumbersnexttothe%;the“run”of
numbersshouldultimatelystopat0.0(+/-0.1).
IfthedigitaldisplaydoesnotreturntoZERO,theunitmostlikelyisinneedofservice.
• Performcalibrationcheck(seeAppendixE).
• ContactBrookeldEngineeringLaboratories,Inc.oryourBrookelddealerforrepair(see
AppendixM).
*Thisproceduredoesnotapplytoinstrumentswithballbearingsuspension(seeSectionI.3).
❏ Display Reading Will Not Stabilize
✓ Specialcharacteristicofsampleuid.Thereisnoproblemwiththerheometer.
• RefertoAppendixC.
✓ Checkforerraticspindlerotation.
• Verifypowersupply
• ContactBrookeldEngineeringLaboratories,Inc.oryourBrookelddealerforrepair.
✓ Bent spindle or spindle coupling.
• ContactBrookeldEngineeringLaboratories,Inc.oryourBrookelddealerforrepair.
✓ Temperatureuctuationinsampleuid.
• Usetemperaturebathforcontrol.