Brookfield DV3T Rheometer Manual do Utilizador Página 36

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 125
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 35
Brookeld Engineering Labs., Inc. Page 36 Manual No. M13-2100
AviolationoftheQCLimitsduringthetestwillbeindicatedbyaashingyellowboxaroundthe
displayforthespeciedparameter.NotethatTimedoesnothaveadedicateddisplay.Aviolation
willalsoresultinanaudiblebeepandaonetimewarningmessage(seeFigureIII-11).Datawill
continuetobecollectedwhilethewarningmessageisdisplayed.
Figure III-11
WhenViscosityorTorquearetheselectedQCLimitsparameter,QCLimitswillbeshownduring
thetestontheRealTimeGraphifthatparameterisselectedfordisplay(seeFigureIII-12).The
QCLimitsarerepresentedbythedashedlines.
TIP:AudiblealarmsmaybeturnedoffintheUserSettings
.
TIP: AnindicationofaQCLimitviolationisnotpartofthedataset.Theusercanrecorda
violationbyusingtheNotesavailablewhenviewingtestResults(seeSectionIII.11).
Trend Bar
Figure III-12
Vista de página 35