Brookeld Engineering Labs., Inc. Page 36 Manual No. M02-215-B0412
III.6 Test Results
III.6.1 Test Passed
When a yield test runs successfully, the resultant Yield Stress (Pa), the % Torque value at the
time of the yield and the Temperature at yield are printed.
Test (#5: Sample Tst) Complete
Yield Stress (Pa) = 726.36 % Torque @ Yield = 73.56 Temperature = 25.5C
In addition, the results are shown on the display as follows:
%=73.56 025.5C
YIELD-726.36 Pa
Figure III-21
III.6.2 Test Failed
Some of the reasons a yield test may fail are as follows.
III.6.2.1 Under-Range Condition
Only yield measurements in the % Torque range from 10% to 100% will be accepted. If
a Yield Stress is reached before the % Torque value is 10%, an Under-Range error causes
the test to end.
The following is sent to an attached printer:
Test (#5: Sample Tst) Complete
Test Failed = Under-range
The following screen is displayed:
YIELD ERROR
UNDER-RANGE
Figure III-22
III.6.2.2 Under-Range Error
The yield stress (Torque Reduction Value) is reached before the absolute % torque reaches
10%. When a test is started, once the absolute torque is greater than zero (0), the second,
third, and fourth Delta Torques (also known as the Base Increment Torque) are averaged.
The rst three (3) Delta Torques after 9% absolute torque is reached are also averaged. An
under-range error occurs if the second average falls below the rst average by 15% or more.