Brookeld Engineering Labs., Inc. Page 47 Manual No. M02-215-B0412
Container size: Use either a 600 mL Low Form Grifn Beaker.
or
the uid container for the viscosity standard.
Temperature: As stated on the uid standard label (±) 0.1°C
Brookeld Viscosity Standard Fluid General Information
We recommend that Brookeld Viscosity Standard Fluids be replaced on an annual basis, one
year from date of initial use. These uids are pure silicone and are not subject to change over
time. However, exposure to outside contaminants through normal use requires replacement on
an annual basis. Contamination may occur by the introduction of solvent, standard of different
viscosity or other foreign material.
Viscosity Standard Fluids may be stored under normal laboratory conditions. Disposal should
be in accordance with state, local, and federal regulations as specied on the material safety
data sheet.
Brookeld Engineering Laboratories does not re-certify Viscosity Standard Fluids. We will is-
sue duplicate copies of the Certicate of Calibration for any uid within two years of purchase
date. Brookeld Viscosity Standard Fluids are reusable provided they are not contaminated.
Normal practice for usage in a 600 mL beaker is to return the material from the beaker back
into the bottle.
Calibration Check Procedure
1) Shut off the rheometer.
2) Place the viscosity standard uid (in the proper container) into the water bath.
3) Press and hold the ↓ and ↑ keys simultaneously while turning the rheometer power ON.
Figure C-1 appears on the rheometer display.
RHEOMETER SETUP
ENTER TO START
Figure C-1
4) Press the ENTER key and Figure C-2 appears on the rheometer display.
CAL VERIFY?
NO↑↓ THEN ENTER
Figure C-2