Brookeld Engineering Labs., Inc. Page 28 Manual No. M03-165-F0612
On entry to the Options menu, the following rules regarding current viscometer operation are in
force:
1. Printer output will be suppressed when in the Custom Speed option, the Time to Torque
and Time to Stop options, the Download A Program and Run A Program options. It
will be continued when any other option is selected.
2. IfthemotorisONwhentheuserenterstheOptionsmenu,choiceswillbelimitedto:
CGS/SI units (under SETUP), °F/°C units (under SETUP), PRINTING SELECTIONS
and PC PROG.
3. The last selectedmenuoptionwillbeashing.
Selecting an Option
The following is a quick reference for entering and using the OPTIONS menu:
Press
OPTIONS
TAB
to enter Options Menu.
Press
or
toscrolltoaspecicoption.
ForOptions:
Press
OPTIONS
TAB
totogglebetweenthechoicesavailableforaspecicoptionwhenindicated.
Press
ENTER
AUTO
RANGE
toselecttheashingoption.
IV.2 Setup
FromthemainOptionsscreen,theuserscrollsupordownuntilthefollowingscreenisdisplayed:
SETUP
DTIME TO TORQUE
Figure IV-2
A press of the ENTER/AUTORANGE key takes you into the Setup sub-menu (Figure IV-3). As in the
main Options menu, you can scroll up or down through the various Setup options. In order to
access all options, the motor must be turned off.
#
$