Brookeld Engineering Labs., Inc. Page 39 Manual No. M03-165-F0612
Activating print selections in the Print mode can only be done by exiting to the main menu and
pressing the
PRINT keyforfour(4)seconds.“P%”willashinfrontofthetorquereading,conrming
that you are now in the Print Interval mode. Pressing
PRINT for one (1) second thereafter will
disable the Print mode and remove the “P%” from the display.
IV.3.4 PC Program (On/O)
Note: ThisoptiondoesnotapplytousewithRHEOCALCsoftware.
This option causes the serial port of the DV-II+Pro viscometer to go into a high speed output
mode(approximately3printlinespersecond)forusewithBrookeldWINGATHER®software
program. When ON, you may enter the Options menu but will not be allowed to make any option
selections until the PC PROG is turned OFF. All front panel keys will function normally when
you turn the option ON and return to normal viscometer operation by pressing the MOTOR ON/
OFF/ESCAPE key. When OFF, the DV-II+Pro will return to the last set print time interval when
printing is resumed.
Note: WINGATHER® software can be used for data acquisition and
analysis. Rheocalc software can be used for complete external
control of DV-II+Pro and data acquisition.
FromtheOptionsmenu,scrolltothescreenshowninFigureIV-32:
SET PRINT TIME
DPC PROG OFF
Figure IV-32
PresstheOPTIONS/TABkeytodisplayFigureIV-33:
SET PRINT TIME
DPC PROG ON
Figure IV-33
Pressing the OPTIONS/TAB key would return you to the screen display of Figure IV-32. Repeated
pressing of the OPTIONS/TAB key would cause you to toggle back-and-forth between the displays
of Figure IV-32 and Figure IV-33.
To turn high speed output ON, press the ENTER/AUTORANGE key when the appropriate screen is
displayed. Then press the MOTOR ON/OFF/ESCAPE key to exit the Setup mode. This returns you
to the default screen display and resumes normal viscometer operation with high speed output
enabled and normal printer operation using the last entered print time interval.