Brookeld Engineering Labs., Inc. Page 66 Manual No. M03-165-F0612
A.2 Setup
1. Be sure that the Viscometer is securely mounted
to the Laboratory Stand, leveled and zeroed
with no cone or cup attached and 0% torque is
displayed.
2. Figure A-2 shows a typical water bath setup.
Connect the sample cup inlet/outlet ports to the
water bath inlet and outlet and set the bath to
thedesiredtesttemperature.Allowsufcient
time for the bath to reach the test temperature.
3. The Viscometer has been supplied with a
special cone spindle(s) which contains the
Electronic Gap Setting feature. The “CPE” or
“CPA” part number designation on the cone
spindle veries the Electronic Gap Setting
feature.
Note: The “CPE” or “CPA” cone or cup
cannot be used with earlier DV-II+
cone/plate Viscometers (below S/
N50969) which do not have the
electronic gap setting feature.
4. With the motor off, thread the cone spindle by
using the spindle wrench to secure the viscom-
eter coupling nut (see Figure A-3); gently push
up on the coupling nut and hold this securely
with the wrench. Thread the cone spindle by
hand.
Note: Left Hand Threads.
5. Attach the cup, taking care not to hit the cone
with the cup (Figure A-4), by positioning the
cup against the mic ring and swinging the ten-
sion bar under the cup. The tension bar should
have plastic tubing in place.
Bath/Circulator
Bath
Inlet
Bath
Outlet
Sample
Cup
Cup
Outlet
Cup
Inlet
Figure A-3
Spindle
Wrench
Cone Spindle
These surfaces
must be clean!
Coupling Nut
Micrometer
Adjustment
Ring
Do Not hit the
CONE with the CUP!
Tension Bar
Figure A-2
Figure A-4