Brookeld Engineering Labs., Inc. Page 82 Manual No. M03-165-F0612
WhenaULorDINULAdapterisused,thewaterbathisstabilizedatthepropertemperature:
1) PuttheproperamountofviscositystandarduidintotheULTube.(RefertotheULAdapter
instruction manual).
2) Attach the spindle (with extension link and coupling nut) onto the DV-II
+Pro.
3) Attach the tube to the mounting channel.
4) Lowerthetubeintothewaterbathreservoir,orifusingtheULA-40Ywaterjacket,connect
the inlet/outlets to the bath external circulating pump.
5) Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test temperature.
6) Measure the viscosity and record the Viscometer reading.
Note: The spindle must rotate at
least ve (5) times before readings are taken.
Calibration Procedure using a Helipath Stand and T-Bar Spindles
WhenaHelipathStandandT-Barspindlesareused:
1) Remove the T-bar spindle and select a standard LV (#1-3) or RV, HA, HB (#1-6) spindle. Follow
theproceduresforLV(#1-3)andRV,HA,HB(#1-6)Brookeldspindlesoutlinedabove.
2) T-Bar spindles should not be used for verifying calibration of the DV-II
+Pro Viscometer.
Calibration Procedure for Spiral Adapter
Brookeldrecommendsatwostepcheck.Firstverifythecalibrationoftheviscometerusingthe
standard viscometer spindles (LV #1-3, RV #2-6, HA #2-6 and HB #2-6 or cone/plate spindles)
as detailed in this appendix. Second verify the calibration of the viscometer using the Spiral
Adapter. The use of an accessory device may increase the accuracy of measurement associated
with the DV-II+Pro.
1) Placetheviscositystandarduid(inthepropercontainer)intothewaterbath.
2) Attachthespindletotheviscometer.Attachchamber(SA-1Y)andclamptotheviscometer.
3) Lower the DV-II
+Pro into measurement position. Operate the viscometer at 50 or 60 RPM
untilthechamberisfullyooded.
4) Theviscositystandarduid,togetherwiththespindle,shouldbeimmersedinthebathfora
minimumof1hour,stirringtheuidperiodically(operateat50or60RPMperiodically),prior
to taking measurements.
5) After1hour,checkthetemperatureoftheviscositystandarduidwithanaccuratethermometer.
6) If the uidisat testtemperature(+/- 0.1°C ofthespecied temperature,normally25°C),
measure the viscosity and record the viscometer reading.
Note: The spindle must rotate at least
ve (5) times for one minute, whichever is greater before readings are taken.
7) TheviscosityreadingshouldequalthecPvalueontheviscosityuidstandardtowithinthe
combined accuracies of the viscometer and the standard (as discussed in the section entitled,
Interpretation of Calibration Test Results). However, instrument accuracy is ±2% of the
maximum viscosity range and not the standard 1%.