Brookeld Engineering Labs., Inc. Page 31 Manual No. M13-167-B0614
Instructions: Recordspecicinstructionstotheoperator.Thisinformationwillbepresented
immediatelywhentheprogramisRun(seeFigureIII-8).Theoperatoris
requiredtoacknowledgethemessagebeforetheprogramwillcontinue.
Figure III-8
TIP:Iftheoperatorselects“Donotshowthismessageagain”within
the Instructions message box, then Instructions will no longer
bedisplayedforanytest.Thisconditioncanberesettoallow
InstructionstobedisplayedintheDisplaysectionofUserSettings
.
TIP:Instructionscanbeviewedduringthetestbyselectinginstructions
(more/lessbar)intheViewTestscreen.
Reports:Denehowdatawillbepresentedattheconclusionofthetestandspecify
thepathforsavingdatales.Upontheconclusionofatest,theusercanbe
presentedwitheitheradataTableoranAverageofcollecteddatapoints.Once
intheResultssection,thedownarrow
canbeusedtochangetheview.
PostTestAveragingcanbespeciedaseitherTestAveragingorStepAveraging.
TestAveragingallowsyoutoselectwhichstepfromtheprogram(createdwith
PGFlash)istobeaveragedanddisplayed.StepAveragingisusedwhenthere
isasingleteststep(ascreatedfromthedisplayofDV2T).SeeSectionIII.10
for more detail relating to Data Averaging.
Thedatapathforsavingdatacanbepreselectedaspartofthetestreport.
TIP: Selectingadatapathcanhelptoorganizedatastorageintheinternal
memoryofDV2T.Bypreselectingthedatapath,operatorsaredirected
toputdatafromspecictestsintospecicareas.
TIP: IfyouselectadatapaththatincludesaUSBFlashDrivelocationandthe
USBFlashDrivehasbeenremoved,anerrormessagewillbedisplayed.
Density:Inputadensityvalueforthesampletobetested.Thisvaluewillbeusedto
calculateviscositywhenkinematicviscosityunitshavebeenchosenfordisplay
(seeSectionIV.4.2).