Brookeld Engineering Labs., Inc. Page 58 Manual No. M13-167-B0614
Appendix A - Cone/Plate Viscometer Set-Up
This Cone/Plate version of the DV2T uses the same operating instruction procedures as
describedinthismanual.However,the“gap”betweentheconeandtheplatemustbeveried/
adjustedbeforemeasurementsaremade.Thisisdonebymovingtheplate(builtintothe
samplecup)uptowardstheconeuntilthepininthecenteroftheconetouchesthesurface
oftheplate,andthenbyseparating(lowering)theplate0.0005inch(0.013mm).
WhenoperatingtheCone/Plateatelevatedtemperature,thegapmustbesetwiththecup
andspindleequilibratedatthetemperaturerecommended.MaximumtemperatureforCone/
Plate operation is 80ºC. Maximum operational temperature of sample cup is 100
ºC. Personal
protectionisrecommendedwhencontrollingtotemperaturesabove80ºC.
Note: MicrometerAdjustmentRingwillbecomehotwhencontrollingsample
cup at temperatures above 50ºC.
ProgrammableDV2TCone/PlateViscometers,S/N50969andhigher,haveanElectronicGap
Settingfeature.Thisfeatureenablestheusertoeasilyndthe0.0005inchgapsettingthat
wasestablishedatBrookeldpriortoshipment.
Brookeldrecommendsthatthemaximumparticlesizeinthesamplematerialformeasurement
withcone/plategeometrybelessthan5timesthegapsetting.Amoreconservativeapproach
istolimitthemaximumparticlesizetolessthan10timesthegapsetting.
ThefollowinginformationexplainshowtosettheElectronicGapandverifycalibrationofthe
DV2TViscometer.
A.1 Electronic Gap Setting Features
TOGGLE SWITCHallowsyoutoenable/disablethe
ElectronicGapSettingFeature:leftpositionis
OFF(disabled),rightpositionisON(enabled).
PILOT LIGHT isthered(LED)light;whenilluminated,
it means the Electronic Setting Function is
sensing(enabled).
Note: Besurethelightisoffbefore
introducingthetestsample.
CONTACT LIGHTistheyellow(LED)light;whenit
rstturnson,the“hitpoint”hasbeenfound.
SLIDING REFERENCE MARKERisusedafterndingthe
“hitpoint;”itisthereferenceforestablishingthe
0.0005inchgap.
MICROMETER ADJUSTMENT RINGisusedtomovethecupupordowninrelationtothecone
spindle.Turningtheringleft(clockwise)lowersthecup;turningitright(counterclockwise)
raisesthecup.Eachlineontheringrepresentsonescaledivisionandisequivalentto
0.0005inchmovementoftheplaterelativetothecone.
Pilot Light
(red)
Toggle Swtich
Contact Light
(yellow)
Sliding Reference
Marker
Micrometer
Adjustment Ring
Figure A-1